Description
Features:– Patented reference design for exceptional performance
– Mounts in Signet standard fitting from 1/2″ to 4″
– ISO 7/1-R 3/4 Thread
– Special design allows for installation at any angle, even inverted or horizontal
– Ryton (PPS) body for broad range of chemical compatibilty
– Patented DryLoc connector with gold plated contacts
– Quick temperature response
– Bulb and flat HF resistant glass available for trace HF, in less than 2% concentration applications
– Low conductivity sensor available for liquids down to 20 uS/cm
Applications:
– Water treatment and water quality monitoring
– Cooling tower and boiler protection
– Neutralization systems
– Sanitization Systems
– Pool and Spa Control
– Aquatic animal life support systems
Specifications
General
– Compatibility: 2750 Sensor Electronics (for 8900, 9900, 4 to 20 mA or Profibus Concentrator
2760 Preamp (8750)
– Tip Design: Bulb
– Low Conductivity applications, 20-100 uS/cm recommended
– Operating Range: pH: 0 to 14 pH
ORP: +/1 2000 mV
– Process Connection: 3/4″, for use in reducing tees up to 4″
– Temperature Sensor: PT1000 Versions: Compatible with Signet 2750 pH/ORP sensor electronics for connection to a PLC or the the Signet 8900 or 9900 instruments
3K Ohm Balco versions: Compatible with the Signet 2760 pH/ORP preamp for connection to the Signet 8750 ph/ORP transmitter also compatible with the 2750 pH/ORP Sensor Electronics
Wetted Materials
– pH: Ryton (PPS), glass, UHMW, PE, FPM
– ORP: Ryton (PPS), glass, UHMW, PE, FPM, Platinum
Max Temperature/Pressure Rating
– Operating Temperature: 14°F to 185°F
– Max Operating Pressure: 100 psi @14 to 150°F, 58 psi @ 150 to 185°F